1.3 RS-232 AND RS-485 JUMPERS SETTINGS
RS232 Interface
Jumper g - installed for normal operation
Jumper h - installed when used as a slave display with
the RS232 output of another meter of the series.
Jumper j - provides a pull-up resistor on the RTS line.
Note: The board is shipped standard with jumpers g
and j installed RS232
RS485
RS485 Interface
h
g
j
a
dc
b
fe
Note: Bias jumpers b and e must be
installed on board for proper operation. If
the outputs of more than 1 RS485 are
connected together, then one (AND ONLY
ONE) RS485 board must have the jumpers
installed.
Full Duplex
Jumpers b and e - bias jumpers installed on
one board
Jumpers a and d - are installed with long cable
runs to add 121 ohm termination resistors. If mul-
tiple meters are on same line, only the last meter
in the line should be jumpered.
Half Duplex
Jumpers b and e - bias jumpers installed on one board
Jumpers c and f - installed for half duplex operation
Jumper a - is installed with long cable runs to add 121 ohm
termination resistor. If multiple meters are on same line, only the
last meter in the line should be jumpered.