-28-
33 Serial Cnfg3 Bit 7 6 5 4 3 2 1 0 Transmit
(Cont'd) 1 0 0 Peak value only
1 0 1 All active items + Peak
0 Termination chars end of all items only
1 Termination chars end of each item
1 Latching RTS (0=Non-Latching RTS)
1 Custom recognition character (0 = “*” recog char)
1 Half Duplex (0 = Full Duplex)
32 Serial Cnfg2 Bit 7 6 5 4 3 2 1 0
X X X X X Counter address 0-31 (5 bits)
1 Command Mode (0 = Continuous)
1 Alarm data included with reading (0 = excluded)
1 LF following CR (0=no LF)
31 Serial Cnfg1 Bit 7 6 5 4 3 2 1 0 Continuous output data rate
0 0 0 0 Reading rate
0 0 0 1 Reading rate / 2
0 0 1 0 Reading rate / 4
0 0 1 1 Reading rate / 8
0 1 0 0 Reading rate /16
0 1 0 1 Reading rate /32
0 1 1 0 Reading rate /64
0 1 1 1 Reading rate /128
Baud rate 1 0 0 0 Reading rate /256
0 0 0 300 baud
0 0 1 600 baud
0 1 0 1200 baud
0 1 1 2400 baud
1 0 0 4800 baud
1 0 1 9600 baud
1 1 0 19200 baud
1 Send filtered value (0 = Send unfiltered value)
30 Options Do not use.This byte determined by option boards installed
2F Filter Bit 7 6 5 4 3 2 1 0 Approximate time constant
0 0 0 No filtering
0 0 1 0.1 sec
0 1 0 0.2 sec
0 1 1 0.4 sec
1 0 0 0.8 sec
1 0 1 1.6 sec
1 1 0 3.2 sec
1 1 1 6.4 sec
1 High adaptive threshold(0=Low threshold)
1 Display filtered input(0=Display unfiltered)
1 Peak val of filtered input(0=Unfiltered input)
1 Conventional filter (0=Adaptive filter)