BF Analog Setup Bit 7 6 5 4 3 2 1 0 bit
0 = Analog Out Unfiltered
1 = Analog Out Filtered
0 = Current Output
1 = Voltage Output
0 0 0 0 0 0 = Not used
35 Decimal point 01 Byte values in hex XXXXX.
02 (2 hex characters/byte) XXXX.X
03 XXX.XX
04 XX.XXX
05 X.XXXX
06 .XXXXX
34 Lockout2 Bit 7 6 5 4 3 2 1 0 bit = 0 is unlocked for all items
1 = SEr 1, Ser 2 locked
1 = An Lo, An Hi locked
1 = Alarm setpoint programming locked
1 = ALSEt locked
1 = Front panel DPM Reset locked
1 = Front panel Peak & Alarm Reset locked
1 = View Alarm setpoints locked
1 = View Peak value locked
33 Lockout1 Bit 7 6 5 4 3 2 1 0 bit = 0 is unlocked for all items
1 = OFFSt locked
1 = SCALE, Lo In, Lo Rd, Hi In, Hi Rd
1 = FILtr locked locked
1 = SEtuP, dEC.Pt locked
1 = InPut locked
0 0 0 = Not used
32 Serial Cnfg2 Bit 7 6 5 4 3 2 1 0
X X X X X = DPM address 0-31 (5 bits)
1 = Command mode (0 = Continuous mode)
1 = Alarm data included with reading (0=excluded)
1 = LF following CR (0= no LF)
31 Serial Cnfg1 Bit 7 6 5 4 3 2 1 0 Continuous Output Data Rate
60 Hz 50 Hz
0 0 0 0 .017s 02s
0 0 0 1 .28 .34
0 0 1 0 .57 .68
0 0 1 1 1.1 1.4
0 1 0 0 2.3 2.7
0 1 0 1 4.5 5.4
0 1 1 0 9.1 10.9