-33-
33 Lockout1 Bit 7 6 5 4 3 2 1 0 bit = 0 is unlocked for all items
1 = Count locked
1 = Setup, config and dEC.Pt locked
1 = Input type locked
1 = Change display item number
1 = Tare Locked
1 = Offset, LO, HI RD locked
1 = View alarm setpoints locked
1 = View peak locked
32 Serial Cnfg2 Bit 7 6 5 4 3 2 1 0
X X X X X = DPM address 0-31 (5 bits)
1 = Command mode (0 = Continuous mode)
1 = Alarm data included with reading (0=excluded)
1 = LF following CR (0= no LF)
31 Serial Cnfg1 Bit 7 6 5 4 3 2 1 0 Continuous Output Data Rate
60 Hz 50 Hz
0 0 0 0 .017s 02s
0 0 0 1 .28 .34
0 0 1 0 .57 .68
0 0 1 1 1.1 1.4
0 1 0 0 2.3 2.7
0 1 0 1 4.5 5.4
0 1 1 0 9.1 10.9
0 1 1 1 18.1 21.8
1 0 0 0 36.3 43.5
1 0 0 1 72.3 86.7
0 0 0 300 baud
0 0 1 600 baud
0 1 0 1200 baud
0 1 1 2400 baud
1 0 0 4800 baud
1 0 1 9600 baud
1 1 0 19200 baud
1 = Send filtered value (0= send unfiltered value)
17 Serial Cnfg3 Bit 7 6 5 4 3 2 1 0 Data Sent
0 0 0 .Net and gross
0 0 1 Net only
0 1 0 Gross only
0 1 1 Peak only
1 0 0 Net, gross and peak
0 CR (LF) after all items (1=after each item)
0 Non-latching RTS (1 = Latching RTS)
0 Not used
0 RS 485 full duplex (1 = Half duplex)
Comentarios a estos manuales