Laurel-electronics LAUREATE SERIES 2 DIGITAL PANEL METER Manual de usuario Pagina 47

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 64
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 46
47
Press Menu
Select Key
Press Digit Select
Key
Press Value Select
Key
_
0
000
U
Line feed
0
No line feed after carriage return
1
Line feed after carriage return
_0
0
00
U
Alarm data with readings
0
No alarm data
1
Alarm data with reading
_00
0
0
U
Control of data output
0
Continuous data output
1
Data output on ASCII command only
.SEr 2.
Serial Setup 2
_000
0U
Meter address with Custom
ASCII protocol
Select
1
thru
F
for addresses 1 thru 15.
Select
0.
thru
F.
(with decimal point) for
addresses 16 thru 31.
0
0000
U
Half or full duplex
0
Full duplex
1
Half duplex
0
0
000
U
Special start & stop char.
0
Standard continuous mode
1
Special start & stop characters
00
0
00
U
RTS mode
0
Normal RTS
1
Single transmission
.SEr 3.
Serial Setup 3
000
0
0
U
Termination characters
0
Only at end of all items
1
At end of each item
.SEr 3.
Serial Setup 3
(continued)
0000
0U
Data sent in continuous
mode
0
Reading
1
Peak
2
Valley
3
Reading + peak
4
Reading + valley
5
Reading + peak + valley
__
0
00
U
Modbus ASCII gap timeout
0
1 sec
1
3 sec
2
5 sec
3
10 sec
__0
0
0
U
Serial protocol
0
Custom ASCII
1
Modbus RTU
2
Modbus ASCII
SEr 4.
Serial Setup 4.
__00
0U
Parity
0
None, 2 or more stop bits
1
Odd, 1 or more stop bits
2
Even, 1 or more stop bits
_Addr
Modbus Address.
Appears only if the
Modbus protocol is
selected.
__
0
00 __0
0
0 __00
0
Select digit to flash.
__
247
Select
0
through
9
for flashing digit.
Address range is 1 to 247.
Vista de pagina 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 63 64

Comentarios a estos manuales

Sin comentarios