16
KEYSTROKES FOR SETUP
If the MENU key does not work, see Section 9 “Enabling & Locking Out Menu Items.”
Press Menu
Select Key
Press Digit
Select Key
Press Value Select
Key
_dC U
DC Volts
__0.2U __2.0U _20.0U 200.0U 600.0U
0.2, 2, 20, 200, 660V FS
_dC A
DC Amps
__2.0a _20.0a _200.0a _5.0a
0.2, 20, 200 mA, 5A FS
_InPut
Selection of signal
input type & range
_rAtio
Strain gauge & ratio
__0.2U __2.0U _20.0U
0.2, 2, 20V FS.
00_00
Display selection with
scale factor of 1.
0 4-1/2 digits (±20,000)
1 Remote display (±99,999)
2 4-1/2 digits, counts by 10 (±20,000)
3 3-1/2 digits (±2,000)
00 _00
Power line frequency
0 Noise minimized for 60 Hz
1 Noise minimized for 50 Hz
00_ 00
Scaling method
0 Scale and offset method
1 Coordinates of 2 points method
2 Reading coordinates of 2 points method
SEtuP
Meter Setup
00_00
Control inputs 1 & 2:
True = logic 1 (0V or
tied to digital ground)
False = logic 0 (5V or
open)
0 1 = Reset, 2 = Meter Hold
1 1 = Function Reset, 2 = Peak or Valley
2 1 = Hold, 2 = Peak or Valley Display
3 1 = Hold, 2 = Tare
4 1 = Peak or Valley Display, 2 = Tare
5 1 = Tare, 2 = Reset
6 1 = 1, 2 = 1, decimal point = XXXXX
1 = 0, 2 = 1, decimal point = XXXX
.
X
1 = 1, 2 = 0, decimal point = XXX
.
XX
1 = 0, 2 = 0, decimal point = XX
.
XXX
7 1 = 1, 2 = 1, decimal point = XXXX
.
X
1 = 0, 2 = 1, decimal point = XXX
.
XX
1 = 1, 2 = 0, decimal point = XX
.
XXX
1 = 0, 2 = 0, decimal point = X
.
XXXX
8 1 = Function Reset, 2 = Display Blank
9 1 = Hold, 2 = Display Blank
A 1 = Peak or Valley, 2 = Display Blank
B 1 = Tare, 2 = Display Blank
C 1 = Valley Display, 2 = Peak Display
D 1 = Tare, 2 = Reset Tare to Zero
Both inputs 1 and 2 set to logic 1 for selec-
tions 2, 4, A, C = Function Reset
Both inputs 1 and 2 set to logic 1 for selec-
tions 0, 1, 3, 5, 8, 9, B, D = Meter Reset