Press Menu
Select Key
Press Digit
Select Key
Press Value Select
Key
00
0
00
Alarm operates at and
above setpoint (active
high) or at and below
setpoint (active low).
0
AL3 active high AL4 active high
1
AL3 active low AL4 active high
2
AL3 disabled AL4 active high
3
AL3 active high AL4 active low
4
AL3 active low AL4 active low
5
AL3 disabled AL4 active low
6
AL3 active high AL4 disabled
7
AL3 active low AL4 disabled
8
AL3 disabled AL4 disabled
000
0
0
Hysteresis mode or
band deviation mode
(see Glossary)
0
AL3 band deviation AL4 band deviation
1
AL3 hysteresis AL4 band deviation
2
AL3 band deviation AL4 split hysteresis
3
AL3 hysteresis AL4 hysteresis
ALS34
Alarm Setup for
relays 3 & 4
(continued)
0000
0
Number of consecutive
readings in alarm zone
to cause an alarm.
0
After 1 reading
4
After 16 readings
1
After 2 readings
5
After 32 readings
2
After 4 readings
6
After 64 readings
3
After 8 readings
7
After 128 reading
dEU1H
Alarm 1
hysteresis
DEU2H
Alarm 2
hysteresis
0
.0000 0.
0
000 0.0
0
00
0.00
0
0 0.000
0
Select digit to flash.
Select
-9
thru
9
for flashing first digit,
0
thru
9
for other flashing digits. Alarms will activate
above the setpoint by the value entered and
deactivate below the setpoint by the value
entered.
DEU1b
Alarm 1 band
deviation
DEU2b
Alarm 2 band
deviation
0
.0000 0.
0
000 0.0
0
00
0.00
0
0 0.000
0
Select digit to flash.
Select
-9
thru
9
for flashing first digit,
0
thru
9
for other flashing digits. Alarms will activate
above and below the setpoint by the value
entered and will deactivate in the middle of
the band.
dEU3H
Alarm 3
hysteresis
DEU4H
Alarm 4
hysteresis
0
.0000 0.
0
000 0.0
0
00
0.00
0
0 0.000
0
Select digit to flash.
Select
-9
thru
9
for flashing first digit,
0
thru
9
for other flashing digits. Alarms will activate
above the setpoint by the value entered and
deactivate below the setpoint by the value
entered.
DEU3b
Alarm 3 band
deviation
DEU4b
Alarm 4 band
deviation
0
.0000 0.
0
000 0.0
0
00
0.00
0
0 0.000
0
Select digit to flash.
Select
-9
thru
9
for flashing first digit,
0
thru
9
for other flashing digits. Alarms will activate
above and below the setpoint by the value
entered and will deactivate in the middle of
the band.