37
Press
Menu
Press Digit
Select Key
Press Value Select
Key
Coordinates of 2 points scaling method if selected under SEtuP
.Lo_In1.
Low signal
input 1.
000000 000000 000000
000000 000000 000000
Select digit to flash.
Select -9 thru 9 for flashing first digit and 0 thru 9
for other flashing digits. Move decimal point location
when flashing.
Lo_ rd1
Reading at
Lo In1.
000000 000000 000000
000000 000000 000000
Select digit to flash.
Select -9 thru 9 for flashing first digit and 0 thru 9
for other flashing digits. Decimal point is fixed by
dEC.Pt1.
.Hi_In2.
High signal
input 2.
000000 000000 000000
000000 000000 000000
Select digit to flash.
Select -9 thru 9 for flashing first digit and 0 thru 9
for other flashing digits. Move decimal point location
when flashing.
.Hi_rd2.
Reading at
Hi In2.
000000 000000 000000
000000 000000 000000
Select digit to flash.
Select -9 thru 9 for flashing first digit and 0 thru 9
for other flashing digits. Decimal point is fixed by
dEC.Pt1.
Special curve offset for square root or custom curve linearization if selected under ConFiG
.rd0_In.
000000 000000 000000
000000 000000 000000
Select digit to flash.
Select -9 thru 9 for flashing first digit and 0 thru 9
for other flashing digits. Decimal point is fixed by
dEC.Pt1. Enter 04.000 for square root extraction
with a 4-20 mA signal.
Scale multiplier
rESoLn.
Resolution
Flashing 6-digit number
in decade steps from
0.00001 to 100000
Press to select. This multiplier R appears with the
Batch mode and can be applied to Grand Total to
set its decimal point.
Quartz crystal time base calibration
_CALib Time base calibration. Do not change. See Calibration section of manual.
Option dependent menu items
SourcE AL SEt. AL S34 dEUn1H dEUn2H dEUn1b dEUn2b dEUn3H DEUn4H DEUn3b
DEUn4b Menu items related to alarm setup These will only appear if a relay board is
detected. If so, please see Section 14.
An_SEt An_Lo An_Hi or An_SEt An_Lo1 An_Hi1 An_Lo2 An_Hi2 Menu items
related to analog output. These will only appear if a single or dual analog output board is
detected. If so, please see Section 15.
_SEr_1 _SEr_2 _SEr_3 _SEr_4 Menu items related to serial communications. These will
only appear if an RS232 or RS485 I/O board is detected. If so, please see Section 16.
Menu lockout items
_Loc_1 _Loc_2 _Loc_3 _Loc_4 Menu items used to enable or lock out (hide) other menu
items. Loc menu items may be locked out by a hardware jumper. Please see Section 9.