-41-
23. DIGITAL INPUTS
Press the
key until
SEtup (Basic meter setup) is
displayed.
Press to display status.
Press again to select
digit. Selected digit will flash.
Digit 5: digital inputs A & B
at J1, Pins 5 and 4.
Press to select value
for flashing digit
0 A: Reset B: Meter Hold
1 A: Function Reset
B: Peak display
2 A: Meter Hold B: Peak
3 A: Meter Hold B: Tare
4 A: Peak B: Tare
5 A: Tare B: Reset
6 External Decimal Pts.1
7 External Decimal Pts.2
1 2 3 4 5
1 2 3 4 5
DIGIT SELECT KEY
VALUE SELECT KEY
MENU KEY
23.1 FUNCTION OF DIGITAL INPUTS
Tare Logical 0 - The present display value is set to zero and stored as an offset value.
Logical 1 - The displayed value is equal to the signal input minus the tare value.
Peak Display Logical 0 - The peak value of the input signal is displayed.
Logical 1 - The present value of the input signal is displayed.
Hold Logical 0 - The meter display and outputs are held at the last reading.
Logical 1 - The display and outputs are updated normally
Reset Logical 0 - The microcomputer reads and resets the meter to nonvolatile memory values
Logical 1 - The meter display and outputs operate normally.
Function Reset Logical 0 - The microC resets peak to present value and resets alarms.
Logical 1 - The meter display and outputs operate normally.
External Decimal Points Input A Input B Decimal Pts 1 Decimal Pts 2
1 1 XXXXX XXXX.X
0 1 XXXX.X XXX.XX
1 0 XXX.XX XX.XXX
00 XX.XXX X.XXXX
23.3 MENU SELECTIONS