29
Press Menu
Select Key
Press Digit Select
Key
Press Value Select
Key
_0000U
Line feed
0 No line feed after carriage return
1 Line feed after carriage return
_0000U
Alarm data with readings
0 No alarm data
1 Alarm data with reading
_0000U
Control of data output
0 Continuous data output
1 Data output on command only
.SEr 2.
Serial Setup 2
_0000U
Meter address with Custom
ASCII protocol
Select 1 thru F for addresses 1 thru 15.
Select 0. thru F. (with decimal point) for
addresses 16 thru 31.
00000U
Half or full duplex
0 Full duplex
1 Half duplex
00000U
Special start & stop char.
0 Standard continuous mode
1 Special start & stop characters
00000U
RTS mode
0 Normal RTS
1 RTS pulse single transmission
00000U
Termination characters
0 Only at end of all items
1 At end of each item
.SEr 3.
Serial Setup 3
00000U
Data sent in continuous
mode
0 Net + gross
1 Net only
2 Gross only
3 Peak only
4 Net + gross + peak
__000U
Modbus ASCII gap timeout
0 1 sec
1 3 sec
2 5 sec
3 10 sec
__000U
Serial protocol
0 Custom ASCII, 8 bits, no parity
1 Modbus RTU, 8 bits
2 Modbus ASCII, 7 bits
SEr 4.
Serial Setup 4.
__000U
Parity
0 None, 2 or more stop bits
1 Odd, 1 or more stop bits
2 Even, 1 or more stop bits
_Addr
Modbus Address.
Appears only if the
Modbus protocol is
selected.
__000 __000 __000
Select digit to flash.
__247
Select 0 through 9 for flashing digit.
Address range is 1 to 247.
Vista de pagina 28