
-
- 7 -
P6 - SIGNAL INPUT DETAIL
2-Wire Process Transmitter
+
-
Excitation return 1
+ Excitation 2
- Signal input 3
+ Signal input 4
DC & Externally Powered Process
-DC
+DC
Strain Gauge
Signal High 1
NC 2
Signal Low 3
Signal High 4
AC & AC+DC True RMS
2, 20, 200, 600V
AC neutral
0.2V, current
NC 1
NC 2
- Signal input 3
+ Signal input 4
Thermocouple
Excitation return 1
- Sense 2
- Signal 3
+ Signal 4
+ Excitation 5
+ Sense 6
Load Cell
Excitation return 1
+ Excitation 2
- Signal input 3
+ Signal input 4
RTD or Resistance - 2 Wire
Excitation return 1
+ Excitation 2
- Signal input 3
+ Signal input 4
Excitation return 1
+ Excitation 2
- Signal input 3
+ Signal input 4
RTD or Resistance - 3 Wire
RTD or Resistance - 4 Wire
For 4-wire load cell connection, jumper
Pin1 to Pin 2, and Pin 5 to Pin 6.
P4 - ANALOG OUTPUT DETAIL
Driving a Load with 4-20 mA
Analog return 1
4-20 mA output 2
500 Ohms max
Driving a Load with 0-10V
Analog GND 1
0-10V output 2
5 kOhms min
Excitation return 1
+ Excitation 2
- Signal input 3
+ Signal input 4
Excitation return 1
+ Excitation 2
- Signal input 3
+ Signal input 4
MOUNTING FOR COOLING
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
Mount transmitters with ventilation holes at top
and bottom. Leave minimum of 6 mm (1/4")
between transmitters, or force air with a fan.
AL 1
AL 2
RESET
POWER
AL 1
AL 2
RESET
POWER
Comentarios a estos manuales