- 17 -
9. DC SIGNAL CONDITIONER BOARD JUMPER SETTINGS
Five voltage and four current ranges are jumper selectable. These also have to be selected in
Instrument Setup software.
Board Revision N
Board Revision P
Voltage Ranges
E1 E2 E3
±200.00 mV
±2.0000 V
±20.000 V
±200.00 V
±600.00 V
A
A
B
B
B
f
f
h
h
g
b
a
b
a
a
Current Ranges
E1 E2 E3
±2.0000 mA
±20.000 mA
±200.00 mA
±5.000 A
A
A
A
A
e, h
d, h
c, h
a, b, h
b
b
b
b
E1
a
E2
b
E3
h
a
g
e
d
c
b
f
B
A
Voltage Ranges Jumpers
FS Input E1 E2
±200.00 mV
±2.0000 V
±20.000 V
±200.00 V
±300V (UL)
±600V (not UL)
A
B
B
B
B
B
f
k
j
h
g
g
Current Ranges Jumpers
FS Input E1 E2
±2.0000 mA
±20.000 mA
±200.00 mA
±5.000 A
A
A
A
A
e, g
d, g
c, g
a, b, g