Laurel-electronics LT SERIES TRANSMITTERS ANALOG INPUT Manual de usuario Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 40
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
- 16 -
9. DC SIGNAL CONDITIONER BOARD JUMPER SETTINGS
Five voltage and four current ranges are jumper selectable. These also have to be selected in
Instrument Setup software.
Board Revisions Q and R
1.
Use 5 mm (0.2") jumpers for locations E1 designated by a capital letter.
2.
Use 2.5 mm (0.1") jumpers for locations E2 and E3 designated by a lower case letter
.
3.
Store spare jumpers on an unused jumper post not associated with a capital letter.
Voltage Ranges Jumpers
FS Input E1 E2 E3
±200.00 mV
±2.0000 V
±20.000 V
±200.00 V
±300V (UL)
±600V (not UL)
A
A
B
B
B
B
f
f
h
h
g
g
b
a
b
a
a
a
Current Ranges Jumpers
FS Input E1 E2 E3
±2.0000 mA
±20.000 mA
±200.00 mA
±5.000 A
A
A
A
A
e, g
d, g
c, g
a, b, g
b
b
b
b
h
g
f
c
e d
b
a
b
a
E1
B
A
E3
E2
Vista de pagina 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 39 40

Comentarios a estos manuales

Sin comentarios