Input Type Jumper Position
E2 E4 E6 E5
Single-ended (signal & return) a, c a, c a, c c
Complementary (differential) b b b c
Complementary (with excitation b b - b, d
and no zero index)
Input Termination Jumper Position
(for differential inputs only) E1 E3 E5
For long input cable runs a a a
For short input cable runs None None None
Note: Do not install E5-a if E5-b and d are installed
Input Phase Jumper Position
E7
Up count, negative transition of B,
A positive ( A leads B ) None
Up count, positive transition of B,
A positive ( B leads A ) a
Count by options Jumper Position
E9
X1 = positive edge of A input None
X2 = positive and negative edges of A input a
X4 = positive & negative edges of both A & B inputs b
13. QUADRATURE SIGNAL CONDITIONER
13.1 SETTING JUMPERS FOR INPUT SIGNAL LEVELS
See "ZERO INDEX SETUP" for jumper locations for zero index signal (Page 35).